Configuration menu
Config. menu
Language
LCD contrast
Invert LCD
Zero
Signal
Active from
Setting options
Language:
en / fr / de / it / da / es /
fi / sv / no / pt
LCD contrast:
1 - 255 (standard 160)
Invert LCD:
Invert (ü) or do not invert
(û)
Zero:
Reset the stored zero value,
see below.
Signal:
Activate (ü) or
deactivate (û)
Active from:
2.0 to 40.0 % (standard
2.0%)
Zero setting
Zeroing should only be performed if zeroing is not enabled for the torque tester in track
mode. In this event, it is possible that the torque sensor will be overloaded.
If the measured value is unstable the torque tester must be repaired. It is recommended
that the torque tester is re-calibrated as soon as possible after zeroing.
Ensure that no torque is applied to the device and remove all tools from the input drive
when zeroing.
Operating instructions
2
Press and hold
screen to display the version screen.
en
160
Press
ü
ü
02.0%
on the measurement
to enter the configuration menu
Select menu item
(e.g. language)
Confirm menu item
Modify setting
(e.g. en to fr)
Confirm setting
Select
and press
to save and quit
Cancel (exit without sav-
ing changes)
25
loading