INT500
Insulation Resistance
Range
400.0kΩ
4.000MΩ
40.00MΩ
400.0MΩ
4.000GΩ
40.0GΩ
200GΩ
* Above specifications only apply when high quality silicone
leads with test clips are being used with no hands touch.
Test Voltage vs. Maximum resistance range:
50V/10.0GΩ, 100V/20.0GΩ, 250V/40.0GΩ, 500V/100GΩ,
1000V/200GΩ
Test Voltage vs. Minimum resistance range:
50V/50.0kΩ, 100V/100.0kΩ, 250V/250.0kΩ, 500V/0.500MΩ,
1000V/1.000MΩ
Short Circuit Test Current: <2mA, +0%, -50%
Test Voltage Accuracy: -0%, +2%+2V
Auto discharge function: discharge time <1 sec for C ≤ 1uF
Maximum Capacitive load: Operable with up to 1uF load
Live Circuit Detection: if ≥ 30V ac/dc at inputs, test inhibited
OL Reading
400.0kΩ
4.000MΩ
40.00MΩ
400.0MΩ
4.000GΩ
40.0GΩ
220GΩ
Resolution
0.1kΩ
0.001MΩ
0.01MΩ
0.1MΩ
0.001GΩ
0.1GΩ
1GΩ
26
Accuracy*
±(2.5%+5D)
±(20%+3D)
EN