Power Probe INT500 Manual Del Usuario página 26

INT500
Frequency
Range
100.0Hz
1000Hz
ACV Minimum Sensitivity: > 60V
Minimum Frequency: 10Hz
Overload Protection: AC/DC 600V
Low Pass Filter
Available for ACV
Add ±4% to specified accuracy @ 50 to 60Hz
Accuracy is specified for 50 to 60Hz
Cut- off Frequency(-3dB): 1kHz
Overload Protection: AC/DC 600V
Capacitor
Range
100.0nF
1000nF
10.00uF
Overload Protection: AC/DC 600V
Earth-bond Resistance (Continuity)
Range
40.00Ω
400.0Ω
4.000kΩ
40.00kΩ
* <1.00Ω add 3dgt
* output current mode in 20mA add 10dgt
* Automatic bi-directional test mode add 0.5%+5D
Output Current: >200mA @ <2Ω or >20mA@ <2Ω
Output Voltage: >4V
OL Reading
100.0Hz
1100Hz
OL Reading
100.0nF
1000nF
11.00uF
OL Reading
40.00Ω
400.0Ω
4.000kΩ
44.00kΩ
Resolution
0.1Hz
1Hz
Resolution
0.1nF
1nF
0.01uF
Resolution
0.01Ω
0.1Ω
0.001kΩ
0.01kΩ
25
Accuracy
±(1.0%+3D)
Accuracy
±(3.0%+10D)
±(3.0%+5D)
Accuracy
±(1.5%+5D)*
±(1.5%+3D)
EN