Power Probe INT500 Manual Del Usuario página 25

INT500
Electrical Specifications
Accuracy is given as ±(% of reading + counts of least
significant digit) at 23°C ± 3°C, with relative humidity Less than
70% R.H., and is specified for 1 year after calibration.
ACV Function
• ACV and ACA specifications are ac coupled, true RMS.
• For square wave, accuracy is unspecified.
• For non-sinusoidal waveforms, Additional Accuracy by Crest
Factor (C.F.): Add 1.0% for C.F. 1.0 to 2.0
• Max. Crest Factor of Input Signal: 3.0 @ 3000 counts
• Frequency Response is specified for sine waveform.
AC Voltage
Range
600.0V
Input Impedance: 10MΩ// less than 100pF
Frequency Response: 50Hz to 400Hz
Overload Protection: AC/DC 600V
DC Voltage
Range
600.0V
Input Impedance: 10MΩ
Overload Protection: AC/DC 600V
Add 2.5% for C.F. 2.0 to 2.5
Add 4.0% for C.F. 2.5 to 3.0
OL Reading
660.0V
OL Reading
660.0V
2.0 @ 4500 counts
1.5 @ 6000 counts
Resolution
0.1V
Resolution
0.1V
24
Accuracy
±(1.5%+5D)(50~60Hz)
±(2.0%+5D)(61~400Hz)
Accuracy
±(1.0%+5D)
EN
loading