3.4.3 Graduation
Set the scale readability.
0.0005, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20.
NOTE: Not all settings are available for each capacity. Refer to the Setup Table 3-3
for available settings.
CAUTION:
Once the graduation value has been changed, the capacity value might
also need to be changed. Please refer to table 3.3 for the corresponding
capacity values and section 3.4.2 on how to set the capacity.

3.4.4 Power On Unit

Set the unit that will be active at power on.
Auto (last unit in use when power was turned off.), kg, g, lb

3.4.5 Zero Range

Set the percentage of scale capacity that may be zeroed.
2% = zero up to 2 percent of capacity
100% = zero up to full capacity
3.4.6 End Setup
Advance to the next menu.
3.5

Readout Menu

Enter this menu to customize display functionality.
3.5.1 Reset
Set the Readout menu to factory default settings.
No
= not reset.
Yes
= reset.
If the Legal for Trade menu item is set to ON, the Stable Range, Averaging Level,
Auto Zero Tracking and Auto Off settings are not reset.
CATAPULT 5000 SERIES SCALES
Reset:
Stable Range:
Filter:
Auto Zero Tracking
Sleep:
Light:
Auto Off
Expand
End Readout
No, Yes
0.5, 1d, 2d, 5d
Lo, Med, Hi
Off, 0.5d, 1d, 3d
Off, On
HI, Med, Low
Off, 1, 5, 10 (min)
Off, On
Exit READOUT menu