3.5.2 Stable Range
Set the stable range.
0.5d
1d
2d
5d
3.5.3 Filter
Set the amount of signal filtering.
LO
= less stability, faster stabilization time (<1 sec.)
mMEd
= normal stability, stabilization time (<2 sec.)
HI
= greater stability, slower stabilization time (<3 sec.)

3.5.4 Auto-Zero Tracking

Set the automatic zero tracking functionality.
OFF
= disabled.
0.5 d
= the display will maintain zero until a drift of 0.5 divisions per
second has been exceeded.
1 d
= the display will maintain zero until a drift of 1 division per second
has been exceeded.
3 d
= the display will maintain zero until a drift of 3 divisions per second
has been exceeded.
NOTE: When the LFT menu item is set to ON, the selections are limited to 0.5d and 3d.
The setting is locked when the hardware lock switch is set to the ON position.
3.5.5 Sleep
Set the terminal sleep functionality.
OFF
= Terminal will not sleep.
ON
= Terminal will sleep.
3.5.6 Light
Set the backlight brightness.
HI
= High brightness.
mMEd
= Medium brightness.
LOwW
= Low brightness.
3.5.7 Auto Off
Set the automatic shut off functionality.
OFF
= disabled.
1
= powers off after 1 minute of no activity.
5
= powers off after 5 minutes of no activity.
10
= powers off after 10 minutes of no activity.