EN-2
STX BALANCES
2.3 Installing the Weighing Pan
Balances with a rectangular pan are placed into the sub-pan as shown and rotated counter-clockwise until it locks. Round
pans are placed straight down on sub-pan.
2.4 Levelling the Instrument
To level the instrument, adjust the feet so the bubble is centered in the circle of the level indicator. Be sure the equipment is
level each time its location is changed.
2.5 Connecting Power
AC Adapter Installation
Connect the AC adapter power cord to the instrument's power input connector, and then connect the AC plug to a suitable
electrical outlet.
Attention: Only use a AC adapter specified by OHAUS.
Option Interface
Security Slot (Anti-Theft lock slot)
Weigh Below Hole
AC Adapter Input Jack
Security Switch
Battery Compartment Cover
Battery Installation
Install the four "AA" batteries with polarity as shown in the battery compartment.
2.6 Initial Calibration
For best results, the instrument's calibration should be checked prior to first use. If adjustment is needed, refer to the
Calibration section of the instruction manual.