Air tubing
Air tubing
Material
ClimateLineAir
Flexible plastic and electrical components
SlimLine
Flexible plastic
Standard
Flexible plastic
Heated air tubing temperature cut-out:  106°F ( 41°C)
Notes:
 The manufacturer reserves the right to change these specifications without notice.
 The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the mask.
 Do not use electrically conductive or antistatic air tubing.
 The temperature and relative humidity settings displayed are not measured values.
Displayed values
Value
Pressure sensor at air outlet:
Mask pressure
Flow derived values:
Leak
Value
1
Pressure measurement
:
2
Mask pressure
1
Flow and flow derived values
:
Flow
2
Leak
1
Results are expressed at ATPD (Ambient Temperature and Pressure, Dry).
2
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.
Pressure accuracy
Maximum static pressure variation at 10 cm H
Without humidification
With humidification
Maximum dynamic pressure variation according to ISO 17510-1:2007
Device without humidification and Standard air tubing / Device with humidification and Standard air tubing
Pressure (cm H
O)
2
4
8
12
16
20
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
Pressure (cm H
O)
2
4
8
12
16
20
18
Range
4–20 cm H
0
2
0–120 L/min
Accuracy
±[0.5 cm H
0 + 4% of measured value]
2
±6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
±12 L/min or 20% of reading, whichever is greater, 0 to 60 L/min
O according to ISO 17510-1:2007
2
Standard air tubing
± 0.5 cm H
O
2
± 0.5 cm H
O
2
10 BPM
15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
10 BPM
15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
Length
Inner diameter
6'6" (2 m)
0.6" (15 mm)
6' (1.8 m)
0.6" (15 mm)
6'6" (2 m)
0.75" (19 mm)
Display resolution
0.1 cm H
O
2
1 L/min
SlimLine air tubing
± 0.5 cm H
O
2
± 0.5 cm H
O
2
20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8