5. INSTALLATION
5.1 SETTING UP THE LIGHTSOURCE
Place the illuminator on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the illuminator may be splashed with liquid.
Absolutely DO NOT use in any environment with explosive or flammable gases.
This illuminator pulls cooling air into its cabinet through a filtered grille in the
bottom panel underneath the front portion and exhausts warm air from the fan
in the rear.
DO NOT block the space in front or below the front of the illuminator or impede
exhaust air flow behind or above the cabinet.
A clear space minimum of 5"/12.7cm behind and above the cabinet is required.
The illuminator should not be placed where its exhaust will influence other
devices, nor where the heated air from other devices will influence the
illuminator.
Ensure the rear power switch is in OFF position.
Ensure that the lamp is properly positioned for use. Open the lamp compartment door. Move the
lamp positioning lever (found to the right of the lamp) clockwise, from horizontal to vertical. As
the positioning cones disengage the lamp should drop about 3mm. Move the lamp positioning
lever counterclockwise to reengage the positioning cones and observe that the lamp rises about
3mm as it returns to operating position. Close the lamp compartment door.
Connect AC power cord to the power inlet located on the rear panel of the illuminator.
Ensure the power cord is completely seated into the power inlet and engage the cord retainer if
so equipped.
Use only power cords provided with the illuminator.
CAUTION
Plug the AC power cord into a wall outlet using the power cord plug supplied with the unit.
To prevent electric shock, connect power cords of peripheral equipment
through medical isolation transformers.
CAUTION
NOTE
When using medical isolation transformer, be sure to check the
ratings.
Ensure that the power cord is connected to the main power with three prong plug (USA use
UL2601-1 rated isolation transformers and/or power strips only).
®
LIT-217 CUDA
Surgical
Rev. D
Page 9 of 107
transformer power
(English)