Important: Please don't continue to press the button more than 30 seconds.
Cleaning and sterilization
Before used in every patient, the cutting tips must be cleaned and sterilized.
Any residues of gutta-percha which still remain on the instrument tip can be removed by
carefully wiping it on a pellet of cotton wool.
When used properly, the cutting tips can be autoclaved no more than 25 times. The
sterilization time depends on the sterilization temperature (at 134°C for 18 minutes).
Troubleshooting
If the cutting tip does not heat up immediately after the operating button is pressed, check
the electrical contacts on the cutting tip.
If these are defective, change the cutting tip for a new one. If the LED on the handpiece
fails to light up during charging, check whether the charging station is connected to the
mains or check the mains supply itself.
8