8. Troubleshooting
If your questions can't be answered with above information, please contact your local agent directly. They might
require a technician to take care of the problem.
Problem
Power is not ON
Low pressure indicator is ON
Patient is bottoming out
Mattress form is loose
No air produced from some air
outlets of the air tube connector
9. Expected service life:
The products are intended to offer safe and reliable operation when use or installed according to the instructions
provided by Apex Medical. Apex Medical recommends that the system be inspected and serviced by authorized
technicians if there are any signs of wear or concerns with device function and indication on products. Otherwise,
service and inspection of the devices generally should not be required.
‧Check if the plug is connected to mains.
‧Check if there is any blown fuse.
‧Check if the connection between quick connector to pump unit is tightly
secured.
‧Check if all tubing connections along mattress are secured.
‧Check if the CPR valve is sealed.
‧Check if there are any air leakage on air cells.
‧Pressure setting might be inadequate for the patient, adjust pressure to
higher level and wait for a few minutes.
‧Check if all the snap buttons or straps of mattress are all securely fastened.
‧Check if the mattress is fixed to the bed frame by straps.
‧This is normal since there is alternating mode. Air outlets take turns to
produce air during their cycle time.
Solution