Press the HOLD key again to enter the next menu. The display will show
Ani.Ft; press the UNIT key to enter the function settings menu. Then press
the ZERO key to select the filtering time of the animal weight among 1, 2,
3, 4 - press the [HOLD] key to confirm the selection.
3.3.3.
Note: Calibration is performed at the factory. Do not recalibrate the scale unless it
is very inaccurate.
Press and hold the ZERO key then press the ON/OFF key. The display will
show CAPU=. Then press the UNIT key to select the maximum load unit in
kg or lb and press the HOLD key to confirm the selection.
Press the HOLD key again to enter the next menu. The display will show
CAP, so press the UNIT key to set the maximum load of the scale. Then
press the UNIT key to move the flashing digit to the right. Press the ZERO
key to increase the value of the flashing digit and press the HOLD key to
confirm your selection.
Press the HOLD key again to enter the next menu. The display shows CALU
=, so press the UNIT key to select the kg or lb calibration unit. Then press
the HOLD key to confirm the selection.
Press the HOLD key again to enter the next menu. The display will show
CAL, so press the UNIT key to set the value of the calibration weight. Then
press the UNIT key to move the flashing digit to the right and press the
ZERO key to increase the value of the flashing digit. Press the HOLD key to
confirm the selection (the value of the adjusting weight should be greater
than 2/3 of full capacity).
Press the HOLD key again to enter the next menu. The display will show
DIV, so press the UNIT key to select the scale. Press the HOLD key to
confirm the selection.
Press the HOLD key again to enter the next menu. The display will show
ADC, so press the UNIT key to select 20. Press the HOLD key to confirm the
selection.
Press the PRINT key. The display shows CAL, then the AD value. Next press
the UNIT key when the stable indicator appears and the display shows a
flashing value of the calibration weight. Place the displayed weight on the
platform. When the display shows a stabilized result, press the UNIT key,
so ----- will appear on the display, the value of AD. Turn off the scale.
Calibration is complete and you can remove the weights.
Turn on the scale, place the tested weight to check if the indications are
correct. If not, repeat the above procedure.
Calibration