Methods
GB
Chlorine HR with Tablet
CLHr
5 – 200 mg/l Cl
Insert the adapter for 16 mm vials.
Fill a clean vial (16 mm Ø) with 8 ml of the water sample and perform
0.0.0
zero calibration (see "Operation").
Add one CHLORINE HR (KI) tablet straight from the foil to the water
sample and crush the tablet using a clean stirring rod.
Add one ACIDIFYING GP tablet straight from the foil to the same
water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablets are dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Zero
Press the [ZERO/TEST] key.
Test
CLHr
The method symbol flashes for approx. 3 seconds.
RESULT
The result is shown in the display in mg/l Chlorine.