Chauvin Arnoux C.A 6549 Manual De Instrucciones página 81

Megaóhmetro
Ocultar thumbs Ver también para C.A 6549:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

After selection of the printing mode :
- If the transmission of data to the printer goes well, the COM symbol flashes at the top left of the
display unit.
- If a problem occurs, the COM symbol remains lit steadily at the top left of the display unit.
6.3.1
IMMEDIATE PRINTING OF THE MEASUREMENT : PRINT RESULT
When this printing mode is selected, the following are printed, in order:
- general information concerning the measurement,
- the measurement result,
- if the T° function was activated, the measurement result referred to the reference temperature,
- for a Timed Run test, the list of recorded samples.
To stop printing, change the setting of the rotary switch.
Depending on the measurement performed, the following models are obtained.
q
Any measurement except srep function measurements :
CHAUVIN ARNOUX C.A 6549
Instrument number: 000 001
Company:...........................
Address:...........................
...................................
Tel.:..................
Fax:...................
Email:.................
Description:.......................
OBJECT:
01
TEST:
INSULATION RESISTANCE TEST
Date
Starting time:
Running time:
Temperature:
Relative humidity:
Test voltage:
Insulation resistance:
-----------------------------------
Rc - calculated resistance 118,5 GOhm
at reference temperature
with ∆T for R/2
-----------------------------------
DAR (1'/30")
PI (10'/1')
DD
Capacitance
-----------------------------------
Elapsed time
Utest
-----------------------------------
00:00:10
1020 V
00:00:30
1020 V
00:00:50
1020 V
...etc...........................................
(printed only in MR mode)
01
31.01.2003
14h55
00:15:30
23°C
.... %
1000 V
385 GOhm
40°C
10°C
1,234
2,345
-,—-
110 nF
(after timed run test)
Resistance
35,94 GOhm
42,0 GOhm
43,5 Gohm
81