Seca 959 Manual De Instrucciones Y Garantía página 84

Ocultar thumbs Ver también para 959:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 180
Malfunction
... if after switch-on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?
... only the item "SYS" is
visible in the rf menu?
... only the items "SYS"
and "lrn" are visible in
the rf menu?
... the items "APrt" and
"Time" are not visible in
the rf menu?
...after calling up the
menu, the "rf" element is
not displayed.
Er:X:11
is displayed?
...
Er:X:12
...
is displayed?
Er:X:16
...
is displayed?
84 •
Cause/Remedy
• The instrument was not able to send any mea-
surement results to the wireless receiver (seca
wireless printer or PC with seca USB wireless
adapter).
– Check that the scales are integrated in
the wireless network.
– Check that the receiver is switched on.
• Nearby RF equipment (e.g. mobile phones) are
interfering with reception.
- Make sure that RF equipment is kept at least
1 metre away from transmitters and receivers
in the seca wireless network.
NOTE:
If such interference is not eliminated, no fur-
ther acoustic warning will be given with sub-
sequent attempts at transmission.
• The wireless module is disabled.
- Activate wireless module (see "Activate
wireless module (SYS)" on page 79).
• The wireless module is activated but no
wireless group set up.
- Set up wireless group (see "Set up wireless
group (Lrn)" on page 79).
• No wireless printer is logged into the wireless
group.
- Use menu item "lrn" to log the wireless printer
into the wireless group (see "Set up wireless
group (Lrn)" on page 79).
• The scale's wireless module is defective.
- Inform the maintenance service
The scales are too high, or too great a load is
applied at one point.
- Reduce load on scales or distribute weight-
more evenly
- Restart scales
The scales have been switched on with too great
a load.
- Reduce load on scales
- Restart scales
Oscillation of the scales has occurred, preventing
determination of the zero point.
- Restart scales
loading

Productos relacionados para Seca 959