Connecting the ultrasound system
Attach the mini‐dock if you use the system with AC‐powered peripherals.
To attach the mini-dock
1 Turn off the ultrasound system.
The mini‐dock may not function properly if you connect the system while the power is on.
2 Place the mini‐dock on the system platform.
3 Slide the mini‐dock feet under the ultrasound system and firmly press the mini‐dock into the
I/O connector on the back of the system.
The mini‐dock is correctly attached when no gap is between the system and mini‐dock.
See Figure
2.
Note: If at any time the ultrasound system should lose connectivity, turn off the system and reattach
the mini‐dock and system.
Always verify that the mini‐dock and system are correctly attached after changing a
transducer. When you change a transducer, the mini‐dock may loosen.
Mini-dock
Figure 2 System platform with system and mini-dock attached
To connect AC power
1 If using a mini‐dock, ensure that it is connected to the system.
2 Connect the cables. (See "Cables and connectors" on page 13.)
• DC power cord (A)
• System AC power cord (K)
System
System latch
H-Universal Stand User Guide