WARNING:
To avoid possible electrical shock or electromagnetic interference, verify proper
operation and compliance with relevant safety standards for all equipment before
clinical use. Connecting additional equipment to the ultrasound system constitutes
configuring a medical system. SonoSite recommends verifying that the system, all
combinations of equipment, and accessories connected to the ultrasound system
comply with JACHO installation requirements and/or safety standards AAMI-ES1,
NFPA 99, IEC 60601-1-1, and IEC 60601-1-2, and are certified according to IEC 60950
(Information Technology Equipment (ITE)).
Equipment safety
Caution:
Excessive bending or twisting of cables can cause a failure or intermittent operation.
Improper cleaning or disinfecting of any part of the accessories and peripherals can
cause permanent damage. For cleaning and disinfecting instructions, see
and disinfecting"
Accessible metal on the signal inputs and outputs on the back and side of the
ultrasound system is not protectively earthed. Do not perform a high current
ground impedance test involving this part.
Clinical safety
WARNING:
To avoid misdiagnosis, do not use non-medical (commercial) grade peripheral
monitors. These monitors have not been verified or validated by SonoSite as being
suitable for diagnosis.
Electromagnetic compatibility (EMC)
See the safety section of the ultrasound system user guide and supplemental guides for 
information on electromagnetic compatibility with IEC 60601‐1‐2, including the 
Manufacturer's Declaration.
on page 9 and the manufacturer's operating instructions.
"Cleaning
H-Universal Stand User Guide
17