5.10 Low Z measurement (elimination of interference
and induced voltages)
Measurement function in 'Low Z' mode eliminates the influence
or interference voltages or induced voltages, making the measure-
ment more accurate and reliable. These voltages may occur due to
capacitive feedback between the live conductors and the unused
conductors located in the vicinity.
To perform the measurement:
set the rotary switch at Low Z,
press MODE
, if an alternating voltage is to be measured,
, if a constant voltage is to be measured,
connect black test lead to COM terminal, and red test lead to
VΩHz%
contact probe blades to measurement points (for DC circuits:
connect red probe to the positive pole),
read the measurement result on the display.
6.1 RANGE button
The button is used to set the measuring range.
To enable the automatic function, press and hold RANGE
button longer than 1 second.
To manually toggle through the measuring ranges, press
RANGE button
This function is not available for measuring the current of
10 A, frequency, duty cycle, diode test, continuity, capaci-
tance and temperature.
button to display the following symbol:
CAP ºC ºF,