6. Commissioning, operation
Fitting a DIN 175301-803 angular connector
1.
Loosen the screw
2.
Loosen the cable gland
3.
Pull the angled socket
4.
Via the mounting hole
5.
Pass the cable with the appropriate cable outer diameter (see
"Connection diagrams") through the cable gland
and the case
.
6.
Connect the cable ends to the connection terminals on the terminal block
in accordance with the pin assignment (see "Connection diagrams" for
the pin assignment).
7.
Press the terminal block
8.
8. Tighten the cable gland
gland and seal are not damaged and that they are assembled correctly in
order to ensure ingress protection.
9.
Place the flat, square gasket
connection pins.
10. Slide the assembled angled socket
transmitter's connection pins.
11. Using the screw
, screw the angled socket to the pressure transmitter,
hand-tight.
WIKA operating instructions pressure transmitter, model S-11
.
.
+
from the instrument.
, lever the terminal block
into the case
.
around the cable. Make sure that the cable
over the pressure transmitter's
+
onto the pressure
out of the case
.
, ring
, sealing
Mounting hole
EN
19