Changing a procedure
 Select the procedure to be changed and press the rotary/push knob.
The "Delete and create" display field will be activated.
 Press the rotary/push knob with the "Delete and create" field selected
to delete the procedure.
New vacuum and rotation values can now be entered to define a new
step in procedure.
Vacuum
 Turn the rotary/push knob to the right/left to select "Vacuum" from
the on-screen options.
 Press the rotary/push knob.
 Turn the rotary/push knob to the right/left to enter a value.
 Press the rotary/push knob.
Duration
 Turn the rotary/push knob to the right/left to select "Duration" from
the on-screen options.
 Press the rotary/push knob.
 Turn the rotary/push knob to the right/left to enter a value.
 Press the rotary/push knob. This will define a new step in the procedure
with the pair of values vacuum-duration and add it to the procedure.
The diagram will be updated.
Rotation
 Turn the rotary/push knob to the right/left to select "Rotation" from
the on-screen options.
 Press the rotary/push knob.
 Turn the rotary/push knob to the right/left to enter a value.
 Press the rotary/push knob.
Duration
 Turn the rotary/push knob to the right/left to select "Duration" from
the on-screen options.
 Press the rotary/push knob.
 Turn the rotary/push knob to the right/left to enter a value.
 Press the rotary/push knob. This will define a new step in the procedure
with the pair of values rotation-duration and add it to the procedure.
The diagram will be updated.
The vacuum-duration and rotation-duration steps can be repeated a ma-
ximum of twenty times.
SET key: Exit menu, store the newly created procedure and activate it as
the current procedure.
ESC key: Exit menu and reset all changes.
Stopping user-defined distillation
The user-defined distillation will stop automatically once all steps in the
currently active procedure have been completed. It is also possible to
stop the process manually by pressing the rotary/push knob with the
"Stop" display field selected.
Stand-by operation
 Press the "Power" button.
 The device will enter stand-by.
 Press the "Power" button again.
 The screen will display "Start" to indicate that the device is ready for
use again.