Service
Turn the rotary/push knob to the right/left to select a menu option.
Press the rotary/push knob.
Turn the rotary/push knob to change the value (the magnitude of the ch-
ange is dependent on the speed at which the knob is turned).
Press the rotary/push knob again to complete the process.
ESC key: Exit menu, all changes to switch settings are reset, the calibration
remains unaffected.
Switching valves and pump
To check functionality for servicing purposes, use the Service menu to
activate/deactivate directly the valves included in delivery or optional and
the pump, e.g. RV 10.5001.
The RV 10.5001 valve switches the water circuit on and off. This row also
shows the current cooling water flow rate.
Calibrate
The "Calibrate" option is used for calibration of the temperature sensor.
The temperature difference displayed will be calibrated to zero. This action
cannot be reversed.
The temperature sensor must be calibrated whenever it is replaced or chan-
ged, or if a new temperature sensor is installed.
Please contact the Service department.
The temperature sensor supplied with the system is calibrated ex-works.
Performing calibration
Activate the "Calibrate" menu option by turning the rotary/push knob.
Remove both sensors. To remove a sensor, push back the outer ring to
release the locking device on the push-in connector, while at the same time
pulling the temperature sensor out of the connector, see fig. 26.
To reconnect a temperature sensor, push it into the push-in connector until
the limit stop is reached; an initial resistance will be felt.
Fill water at room ambient temperature into a glass beaker (approx. 500
ml). Fully submerge both sensors into the water, see fig. 27.
Wait until the temperature display in the "Service" menu, menu option
"Calibrate" has stabilised, e.g. 0.2K.
Press the rotary/push knob.
The temperature difference displayed will be calibrated to zero.
Calibration is only possible within the temperature range +0.5 K to -0.5 K.
If the temperature is outside this range, then either a temperature sensor
is faulty or connected incorrectly. Please contact the Service department.
If the characteristics of the cooling water supply deviate significantly from
the specifications given in the "Technical specifications", with the result
that the initial conditions for automatic distillation are not achieved, then
it may be necessary to recalibrate the temperature sensors. This may,
however, lead to a reduction in the accuracy with which the distillate
quantity is measured in "Auto-distillation" mode.
Operation hours
This value serves as informaton and cannot be changed.