Using the Oscilloscope
Adjust the vertical sensitivity
You can increase or decrease the vertical sensitivity (Volt/div) of a waveform by
pressing
Channel coupling
Toggle
AC coupling blocks any DC component in the waveform and allows only the AC
component of the signal to be viewed.
DC coupling allows both AC and DC components to pass into the handheld scope.
Probe setting
Press
Toggle
probe. Channel sensitivity and offset, trigger level, measurement results, and
math functions will reflect the unit you have selected.
Press
measuring voltage/current values with a voltage/current probe. The attenuation
factor/sensitivity must be set according to the probe being used to ensure the
measurement results reflect the actual voltage/current level.
46
.
Vertical sensitivity
to set the channel coupling.
Coupling
to access the probe setup submenu.
to set the Volts unit for a voltage probe or Amps for a current
repeatedly to set the attenuation factor/sensitivity for
Keysight U1610/20A User's Guide