Integra MAYFIELD Infinity XR2 Manual De Instrucciones página 6

Ocultar thumbs Ver también para MAYFIELD Infinity XR2:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
EN – ENGLISH
Cleaning of the MAYFIELD Infinity XR2 Tri-Star Swivel Adaptor
Do Not Steam Sterilize! The Plastic components may be damaged by heat
Following steps are recommended:
1. Remove Infinity XR2 Tri-Star Swivel Adaptor from base unit link arm.
2. The Infinity XR2 Tri-Star Swivel Adaptor should be thoroughly cleaned after each use. Scrub
component with a soft brush and use only a pH-neutral detergent. Clean thoroughly to remove
any traces of blood and/or debris and to prevent such blood or debris from interfering with
function or movement. Rinse thoroughly with clean water.
3. Dry all components with a soft dry towel.
4. After component is totally dry, it is ready for storage or next use.
Optional Automatic Wash Cycle:
Any deviation from this guideline could result in damage to equipment as well as
improper cleaning results. Parameters given below must be followed exactly in order
to assure proper cleaning and avoid damage to the equipment.
• Remove the device as indicated in section Cleaning of the MAYFIELD Infinity XR2 Tri-Star
Swivel Adaptor.
• Rinse components under warm tap water prior to placing in washer.
• Arrange device in a way to avoid contact between components.
• Select instrument cycle and ensure proper programming.
WARNING!
CAUTION!
4
loading

Este manual también es adecuado para:

A2111