Audible Continuity - Keysight U1251B Guia Del Usuario

Ocultar thumbs Ver también para U1251B:
Tabla de contenido

Audible continuity

In the range of 0–500 , the beeper will sound if the resistance value falls below
10 . For other ranges, the beeper will sound if the resistance falls below the
typical values indicated in
Table 2-3
Audible continuity measurement range
Measurement range
5.0000 k
50.000 k
500.00 k
5.0000 M
50.000 M
500.00 M
Conductance
Set up the multimeter to measure conductance as shown in
test points and read the display.
The conductance measurement enables the measurement of very high resistance
of up to 100 G.
As the high-resistance readings are susceptible to noise, you can capture the
average readings by using the Dynamic Recording mode. Refer to the section
"Dynamic Recording"
Keysight U1251B and U1252B User's and Service Guide
500.00 
Table 2-3
below.
on page 78 for more information.
Making Measurements
Beeper sound threshold
< 10 
< 100 
< 1 k
< 10 k
< 100 k
< 1 M
< 10 M
Figure
2
2-9. Probe the
63
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

U1252b

Tabla de contenido