Steinberg Systems SBS-PF-600/100N Manual De Instrucciones página 36

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 108
ENTER to confirm the last digit and set the next one. If the digit is the last one, press
ENTER to set the upper limit.
2. Setting the lower limit - after starting the setting, the set lower limit will be shown
and the "-LL-" signal will be displayed alternately. The operation is the same as for
the upper limit setting. After completing the setting of the lower limit, the scale
proceeds to setting the alarm method.
3. Setting the alarm method :
When the scale displays "-IN-", it means an alarm within the limit.
When the scale displays "-OUT-", it indicates an alarm out of the limit.
When the scale displays "-NO-", it means no alarm.
You can press ↑ to switch. Press ENTER to confirm.
Press SET to exit the setting state during setting. The value set will remain blank.
The alarm will be signaled by warning lights - if the value exceeds the upper limit, a
red light will turn on; if the value is below the lower limit, a yellow light will turn on;
if the value is between the lower and upper limit, the green light will turn on.
3.3.5. Calibration
Press and hold the CALI button for approximately 4 seconds. Calibration is
performed after displaying the word "CAL." on the display.
There are two ways to calibrate:
Full load capacity: Place a sufficient amount of load on the platform,
according to the full range of the scale. Press ENTER to confirm. Calibration
complete.
Other load:
o
o
3.4.
Cleaning and maintenance
a)
Pull out the mains plug before each cleaning and when the unit is not in use.
b)
Use only non-corrosive cleaning agents for cleaning the surfaces.
c)
Store the unit in a dry and cool place protected from moisture and direct
sunlight.
d)
Do not spray the unit with a stream of water or immerse it in water.
e)
Make sure that no water enters through the ventilation openings in the casing.
f)
Clean the ventilation openings with a brush and compressed air.
g)
Perform regular inspections of the unit checking technical fitness and any
damages.
h)
Do not leave the batteries in the unit when it will not be used for a long time.
Press SET, the scale window displays the scale that was calibrated
last time.
Press ↑ and ENTER to enter the value of the load to be loaded on
the scale and then put this load on the scale. Press ENTER to
confirm. Calibration complete.
loading