Multimeter Features

Making Scale Transfers (Scale)

The Scale operation emulates a transducer by helping you to convert the
measured readings proportionally to the specified ratio and unit display. Use
Scale to transfer voltage readings to proportional readings when using clamp-on
current probes, high voltage probes, or temperature auxiliary probes. The
available scale conversions are shown in the table below.
Table 3-1
Available scale conversions
Scale item
[b]
1000 V/V
[c]
1 °C/mV
or
[c]
1 °F/mV
1 A/mV
0.1 A/mV
0.01 A/mV
1 mA/ mV
0.1 mA/ mV
[a] The transfer formula used is: Display = Multiplier × Measurement
[b] The scale item is selected from the Setup menu. See
[c] Dependent on temperature unit setup.
If °C or °C°F is selected, 1 °C/mV is shown as the selected scale item. If °F or °F°C is selected, 1 °F/mV is shown as the selected scale item
instead.
96
[a]
Unit
Mul tiplier
1000.0
V
°C
1000.0
or
°F
1000.0
A
100.0
A
10.0
A
1.0
A
0.1
A
"Changing the scale conversion value"
Best resolution
0.1 V
0.1 °C
or
0.1 °F
0.1 A
0.01 A
0.001 A
0.1 mA
0.01 mA
on page 116 for more information.
Keysight U1231A, U1232A, and U1233A User's Guide
Start range
600.0 V
600.0 °C
or
600.0 °F
600.0 A
60.00 A
6.000 A
600.0 mA
60.00 mA