Testing for Continuity
Set up your multimeter to test for continuity as shown in
test points and read the display.
To avoid possible damage to your multimeter or to the equipment under
CAUTION
test, disconnect the circuit power and discharge all high-voltage capacitors
before testing for continuity.
Continuity is the presence of a complete path for current flow. The continuity
NOTE
test features a beeper that sounds and a backlight that flashes as long as a
circuit is complete if short continuity is selected (or broken if open continuity is
selected). The audible and visual alert allows you to perform quick continuity
tests without having to watch the display.
In continuity, a short means a measured value is less that the threshold resistance
values listed in
Table 2-1
Threshold resistance values
Measuring range
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
60.00 MΩ
Press
or open continuity test (
Keysight U1231A, U1232A, and U1233A User's Guide
Table
2-1.
to switch between resistance measurement, short continuity test ( ),
). See
Threshold resistance
<23 ±10 Ω
<230 ±100 Ω
<2.3 ± 1 kΩ
<23 ± 10 kΩ
<131 ± 60 kΩ
<131 ± 60 kΩ
Figure 2-10
to learn more.
Making Measurements
Figure
2-11. Probe the
2
61