Enable The Ac/Dc Mv Measurement - Keysight Technologies U1231A Guia Del Usuario

Ocultar thumbs Ver también para U1231A:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

4
Multimeter Setup Options

Enable the AC/DC mV measurement

You can choose to set the multimeter to measure AC or DC mV at the rotary
positions shown below.
– U1233A:
– U1232A:
– U1231A:
Use this Setup item to enable AC/DC mV measurements. You are recommended
to use the AC/DC mV measurements to precisely measure low voltages.
Parameter
mV
– When this Setup item is enabled, the original functions of the rotary switch
NOTE
– For AC/DC mV measurements, the measurement range is fixed at 600 mV
– Press
Figure 4-15
118
AUX
Hz
AUX
Range
on or oFF
positions shown above are disabled and replaced by AC/DC mV
measurements.
and the input impedance is typically 10 MΩ.
to switch between DC mV, AC mV, and frequency measurements.
P ress
Press
Enable the AC/DC mV measurement
Keysight U1231A, U1232A, and U1233A User's Guide
Defaul t setting
oFF