6.1 Button HOLD/REL
6.1.1 HOLD function
This function is used to 'freeze' the measurement result on the
display. To do this, shortly press HOLD/REL button. When the func-
tion is enabled, the display shows symbol HOLD.
To return to the normal operation mode of the device, press
HOLD/REL button again.
6.1.2 REL function
This mode enables a measurement relative to a reference value.
To enable the mode, press and hold HOLD/REL. Then, the dis-
played readout value is taken as the reference value, and the
readout will be reset.
From this moment, the readings will be presented as the ratio of
the measured value to the reference value.
To enable the mode, press HOLD/REL.
The displayed main result is the difference between the reference
value (read-out at the moment of activating REL mode) and the cur-
rent read-out. Example: if the reference value is 20 A, and the cur-
rent reading is 12.5 A, then the main result on the display will be
-7.5 A. If the new reading is identical to the reference value, then
the result will be zero.
When the function is activated, the automatic adjustment
of the measuring range is not available.
If the reading is outside the measurement range, symbol
OL is displayed In this situation, turn off the function and
manually switch to a higher range.
This function is not available for diode test.