4.4.2. Adjust a Calibrated Temperature

1.
To adjust a calibrated temperature, the unit must first be in the top level of the
settings menu. Please refer to Accessing / Exiting Section for how to enter
the top level menu. Then wait for the screen to show CAL
2.
Briefly press the left knob to enter the Calibration settings menu.
Briefly press the left knob to enter the SPC settings menu.
3.
4.
Rotate the left knob to select which calibration temperature to adjust if two
calibration records are stored. The unit can store1 point (Plate) and 1 point
(Probe) single point calibration record. For calibration performed with an
external probe, the screen will display
5.
Press and hold the left knob until the unit beeps and the
begin calibration. Then repeat steps in the previous Start a Single Point
Calibration Section from step 6 to the end to finish the process.
TM
Guardian
.
.
Hotplate-Stirrers
illuminates to