Cleaning / Checking the Motor
6
6 – 1 Cleaning
6 – 1 – 1 Manual Cleaning
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
For products with moving parts, perform cleaning while moving all parts (see "2 Moving Parts of Each Product").
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
_ Cleaning
1
Remove the dirt on the surface of the product using a soft
bristle brush moistened with neutral enzymatic detergent.
≥ 1 min
Water quality: Distilled, deionized or demineralized water.
* Detergent used for validation:
Neutral enzymatic detergent: neodisher MediZym (DR.
WEIGERT)
_ Drying
1
Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
6 – 1 – 2 Automated Cleaning
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
For products with moving parts, perform cleaning while moving all parts (see "2 Moving Parts of Each Product").
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
2
Rinse the product under running tap water.
For products with moving parts, perform cleaning while moving
all parts (see "2 Moving Parts of Each Product").
≤ 38°C, ≥ 3.5 L/min, ≥ 10 sec.
Water quality: Drinking water
3
Rinse the product surface with distilled, deionized or
demineralized water.
2
Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Checking (see "6 – 2 Checking").