Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
5
With the product immersed in the cleaning tank, pull the syringe
plunger to suck and discharge the cleaning solution. Perform
this step 5 times to retain the cleaning solution inside the
product.
Immerse the product in the cleaning tank for 20 - 30 min.
Push and pull the syringe plunger to discharge the cleaning
solution remaining in the product.
6
Immerse the product in the soak tank containing tap water, then
operate the syringe to suck and discharge water 10 times to
rinse inside the product.
≤ 38°C
Water quality: Drinking water
_ Drying
1
Screw the Cleaning Adaptor G into each Cleaning Adaptor.
2
Push the air gun into the Cleaning Adaptor G and remove
moisture inside the product.
Allowable air pressure
Purging Time
x5
x10
< 500 kPa
≥ 5 sec.
12
7
Rinse the product under running tap water.
For products with moving parts, perform cleaning while moving
all parts (see "2 Moving Parts of Each Product").
≤ 38°C, ≥ 3.5 L/min, ≥ 10 sec.
Water quality: Drinking water
8
Put distilled, deionized or demineralized water into the syringe,
screw the syringe into the product, and discharge water from
the product tip.
Perform this step 10 times.
9
Rinse the product surface with distilled, deionized or
demineralized water.
3
Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
4
Remove the Cleaning Adaptor from the Attachment by
depressing the Release button.
5
Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Lubrication (see "7 – 2 Lubrication")
x10