6. Operation

6.5 HOLD function

HOLD function can be used to freeze display number. In this mode, the scale can catch a dynamic
f
number, hold a stable number, or average a unstable number, then HOLD (freeze) this number
temporary for user to watch or record. This function can be used in normal weighing mode, counting
mode and percent weighing mode. The speed of A/D converter is increased to 80Hz (if CON-
FIG-AD.H.SPD is set to YES) from original 10Hz for some dynamic weighing applications. With the
hold function, it is possible to weigh restless weighing samples such as live animals, moving objects.
The indicator provides special mode settings to accommodate sample's movements.
To make HOLD function be active, the CONFIG-FUNC-HOLD menu item must be set to YES; menu
f
items of USER-HOLD-HLD.MOD /-AVG.TIM /-HLD.TIM /-DYN.RNG /-STB.TIM, USER-OTHER -NLD.
RNG need to be set at a reasonable value. To speedup sampling of weight, set CONFIG-AD.H.SPD
menu item to YES. Please note: lower speed will get more stable data, but some instant weight
information may be lost. To enter HOLD working mode, press [HOLD/SETUP] key when scale works in
normal weighing mode, counting mode or percent weighing mode.
There are several HOLD mode to freeze display data:
f
(1) Positive Peak Number HOLD mode
(2) Negative Peak Number HOLD mode
(3) Toggle HOLD mode
(4) Average HOLD mode
(5) Auto HOLD mode
Positive Peak HOLD:
f
When USER-HOLD-HLD.MOD is set to PS.PEAK, the hold mode is positive peak hold mode. When
scale first enters this working mode, it will display the largest positive number that is from the time of
zero-point set. After entering this working mode, scale will always catches and refresh positive larger
number and display it. To exit HOLD mode, press [HOLD/SETUP] key again.
Negative Peak HOLD:
f
When USER-HOLD-HLD.MOD is set to NG.PEAK, the hold mode is negative peak hold mode. When
scale first enters this working mode, it will display the largest negative number that is from the time
of zero-point set. After entering this working mode, scale will always catches negative larger number
and display it. To exit HOLD mode, press [HOLD/SETUP] key again.
Toggle HOLD:
f
When USER-HOLD-HLD.MOD is set to TOGGLE, the hold mode is toggle hold mode, a manual Hold
function. After entering this working mode, the scale will freeze and display the number if scale is
stable. Only the weight that is over USER-OTHER-NLD.RNG (zero ‚dead' band ) can be held. To exit
HOLD mode, press [HOLD/SETUP] key again. If the time of scale being unstable is more than
USER-HOLD-STB.TIM, STB.ER will be shown, press [TARE/PRESET] key to start averaging again or
press [HOLD/SETUP] key to exit.