Guidance and manufacturer's declaration - electromagnetic immunity
The Phyaction-series devices are intended for use in the electromagnetic
environment specified below. The customer or the user of a Phyaction-series
device should assure that it is used in such an environment.
Immunity
IEC 60601
test
test level
Conducted RF
3 V
rms
IEC 61000-4-6
kHz 80%
150 kHz to 80
MHz
Radiated RF
3 V/m AM 1 kHz
IEC 61000-4-3
80%
80 MHz to 2,5
GHz
Radiated RF
3 V/m CW 200
ENV 50204
Hz d.c. 50%
895 MHz to 905
MHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
The guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey can be considered. If the
measured field strength in the location in which a Phyaction-series device is used exceeds the
applicable RF compliance level above, the Phyaction-series devices should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the Phyaction-series device.
b Over the frequency range 150 kHz to 80 MHz, field strengths must be less than 10 V/m.
Phyaction U
Compliance level Electromagnetic environment
AM 1
10 V......0,15-80 Mhz
51 V...........6,78 Mhz
54 V.........13,56 Mhz
50 V.........27,12 Mhz
45 V.........40,68 Mhz
10 V/m..0,08-1,0 Ghz
26 V/m....1,4-2,0 Ghz
30 V/m...433,92 Mhz
30 V/m........915 Mhz
30 V/m.895-905 Mhz
47
- guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of a
Phyaction-series device, including
cables, than the recommended
separation distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 0,35
p
d = 0,07
p
d = 0,06
p
d = 0,07
p
d = 0,08
p
d = 0,35
p
80 MHz to 800 MHz
d = 0,70
p
800 MHz to 2,5 GHz
d = 0,12
p
d = 0,23
p
d = 0,23
p
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
be less than the compliance level in
each frequency range
Interference may occur in the
vicinity of equipment marked with
the following symbol:
a
, should
b
.