PAGE 12 OF 121
10. Trouble Shooting
P
ROBLEM
Possible Cause
Improper connections
Bulb beyond useful life of
500 hours
Main line fuse blown
(no power to unit)
During the course of normal use per the instructions in this IFU, wear and tear can occur. Below is a list
of observable indications of normal wear and tear, that signify the device should be sent to IlluminOss
for service.
•
Console
o Damage to the power cord
o Tongs misshapen
o Dents
o Dings
o Loose fittings
o Fuse housing / power inlet broken or damaged
o Fuse housing missing
o Loose or missing screws
o Fan Filters are clogged / damaged/ missing / degraded
o Red "Change Bulb" indicator light flashing
o Red "Change Bulb" indicator light steadily illuminated
o Continuous audible beeping sound when device is on, indicating bulb needs to be changed.
•
Light Guide
o Malformed
o Pinched or kinked
o Cracked component
900093.W issued 01-JUL-21
WARNING: ONLY QUALIFIED MAINTENANCE PERSONNEL SHOULD ATTEMPT THE
FOLLOWING PROCEDURES.
: B
W
ULB
ILL NOT IGNITE OR
Visually inspect all input /
(i.e. power cord, bulb)
Replace with a new bulb /
Remove fuse from power
W
HEN TO
PHOTODYNAMIC LIGHT CURING SYSTEM INSTRUCTIONS FOR USE
C
HANGE
Testing
output connections
reflector assembly
receptacle and check
with an ohmmeter
R
S
EQUEST
ERVICE
B
ULB LIGHT IS ON
Corrective Action
Secure all connections
Contact IlluminOss Medical
or your distributor for
bulb replacement
Replace fuse if defective
-
ILLUMINOSS MEDICAL INC