Ensure that the LED display shows 2000 seconds,
then insert the Variable Timer Key contained in the
package of the catheter being used in the procedure.
Confirm that the LED display changes within 2
seconds after insertion and displays the correct time
value corresponding to the key inserted (indicated by
the time value printed on key card, and displayed on
original device packaging).
CAUTION FAILURE TO USE THE PROPER VARIATBLE TIMER KEY CAN RESULT IN INADEQUATE CURING OF THE
ILLUMINOSS MONOMER. ENSURE YOU ARE USNG THE CORRECT TIME KEY CORRESPONDING TO THE IMPLANT
SELECTED FOR SURGERY.
NOTE: there is a protective shroud protruding from the faceplate which prevents improper insertion of the
Variable Timer Keys. See image below for help with insertion.
Ensure that the STANDBY TOGGLE SWITCH is in the
"Foot Pedal Enabled" position, then operate the
Light Source Console by depressing the Foot Pedal.
The Light Source Console will operate for the
prescribed number of seconds. With the STANDBY
TOGGLE SWITCH in the "System Standby Mode," the
Foot Pedal is inoperable and the user will hear an
audible beep when the foot pedal is depressed.
CAUTION: DO NOT REMOVE TIMER KEY DURING COUNTDOWN!
CAUTION: If STANDBY TOGGLE SWITCH is changed to "System Standby Mode" during operation,
wait for time cycle to expire, remove key and reinsert, then reinitiate full cure cycle.
CAUTION: This is an arc, not a filament bulb. Avoid turning the console off and on. It is
recommended that the console is turned on at the beginning of the procedure and remains in
operation until the end of the procedure.
CAUTION: The bulb must be cool before it can be reignited. Turn the unit off and allow 5 – 10
minutes for it to cool down. If the bulb fails to reignite, refer to the "Troubleshooting" section
of this manual. Bulb life is reduced each time the unit is switched on and off. Avoid repeated
cycles that shorten bulb life by leaving the unit on throughout the procedure.
900093.W issued 01-JUL-21
PHOTODYNAMIC LIGHT CURING SYSTEM INSTRUCTIONS FOR USE