2.3 Technical data

2.3.1 Product-specific data
Jar sizes:

2.3.2 Dimensions and weight

TOPITEC
EXPERT (device dimensions) including touch screen
®
Width:
Depth:
Height:
Total weight:
2.3.3 Power supply / electrical parameters
Operating voltage:
Phases:
Protective earth conductor: PE (yellow/green)
Frequency:
Protection rating:
Overvoltage category:
EMC:
Protection class:
Installation standard:

Power supply

Fuses:
Conductor cross-section:
Rated power:
68
TOPITEC
®
EXPERT User Manual | 11-2018
Jars up to max. diameter 118 mm
Single-patient containers, single-patient jars,
jars for stock preparation
approx. 286 mm
approx. 377 mm
approx. 733.5 mm
approx. 19 kg
1 x 230 V
AC
±10%
1 Ph / PE
50 Hz ± 1%
I
2
Laboratories, domestic environments, SMEs
IP 20
DIN EN 61010-1: 2011
5 x 20 medium-acting 10A (A-D14-1845)
min. 1.5 mm
approx. 1800 W
Warning
The device must be connected to a safety socket. The mixing system is
equipped with a fine protection fuse. Medium and general protection must
be provided as part of the power distribution system within the building.
The device is protected with two fuses
(A-D14-1845 5 x 20 medium-acting 10A) installed at the two-pole power
switch. At the location of operation, it must be protected by a 10A
circuit breaker (characteristic B). When replacing a fuse, use one of the same
type, rating and characteristic.