4.
Commissioning

4.1 Setup instructions

Protect the device from shocks, vibrations, strong heat or cold, drafts, chemicals and moisture by
choosing a location where it will be installed.
Place the scale on a firm, free and horizontal surface.
f
Make sure that no cables or other objects are trapped under the scale.
f

4.2 Power supply

Power is supplied as standard via a 6V 2.3Ah rechargeable lead-acid battery or the integrated mains
cable.
There are three battery indicators on the display:
High = The battery voltage is greater than 6.3V
Medium = The battery voltage is between 6V and 6.3V
Low = The battery voltage is lower than 6V
Note:
If the battery voltage falls below 5.6V, the balance will automatically switch off.
When „-bAtLo-" is displayed, the battery needs to be charged. Connect the power cord to charge the
battery. The full charging time is approximately 10-12 hours. The balance can continue to operate on
AC power while charging is in progress.
Battery life and recharge time vary depending on use. Over time, the operating time per full charge will
decrease. If the operating time is no longer acceptable, the battery must be replaced. If the balance is
stored for an extended period of time, the battery must be recharged every 90 days to prevent
premature deterioration in performance.
To save battery power, the scale has two energy saving functions:
Power save mode: The display will darken after 40 seconds if there is no weight on the platform,
f
but the scale is not turned off. When you put weight on the platform, the display will automatically
turn on again.
Automatic switch-off: The scale will turn off automatically after 10 minutes if no key is pressed or
f
the weight is changed.