1.1.3. During use

Never exceed the protective limiting values given in the characteristics specific to
each type of measurement.
When the multimeter is linked to measurement circuits, do not touch unused
terminals.
When the scale of the value to be measured is unknown, check that the scale
initially set on the multimeter is the highest possible or, wherever possible, choose
the autoranging mode.
Before changing functions, disconnect the test leads from the circuit under test.
When performing current measurements, never change of range, do not connect or
disconnect leads without first isolating the current. If you do, there is a risk of
generating surge currents which can blow the fuse or damage the instrument.
Never perform resistance measurements on live circuits.
1.1.4. Symbols
Warning: Risk of danger. Refer to the operating manual to find out the nature of the
potential hazards and the action necessary to avoid such hazards.
Earth
1.1.5. Definition of measurement categories
CAT III: Measurement category III corresponds to measurements on building
installations.
Example: measurements on distribution panels, cabling, etc.
1.1.6. Opening the instrument (performed outside explosive atmosphere only)
Before opening the instrument, always disconnect from all sources of electric current
and make sure not to be loaded with static electricity, which may destroy internal
components.
The F1 fuse certified for EX classified danger zones, as well as the F2 fuse, must be
replaced by a model identical to the origin fuses.
(See §. 3.5.3. Replacing the battery or one fuse).
Any adjustment, maintenance or repair work carried out on the multimeter should be
carried out only by appropriately qualified personnel, after having taken into account
the instructions in this present manual. A "qualified person" is one who is familiar
with the installation, construction and operation of the equipment and the hazards
involved. This qualified person must be familiar with equipment for explosive
atmospheres and be trained in the general rules for intrinsic safety.
When the instrument is open, for maintenance purposes for example, remember
that some internal capacitors can retain a dangerous potential even after the
instrument is powered down.
If any faults or abnormalities are observed, take the instrument out of service and
ensure that it cannot be used until it has been checked out.
It is recommended to remove the battery from the instrument if not used.
26
Dual insulation
According to WEEE directive 2002/96/EC
Intrinsically safe portable digital multimeter
Chapter II