Span [SpaN]
Initiates a span calibration procedure (zero and span)
Lin [LIN]
Initiates a linearity calibration procedure (zero, mid-point and span).
End Cal [End]
Advance to the next menu or return to the top of the current menu.

4.3 Setup Menu

Enter this menu to set balance parameters.
Reset:
Filter:
Auto Zero Tracking:
Stable:
Backlight:
Auto Tare:
Auto Off:
End Setup:
Note: Bold always represents factory default value.
Reset [reset]
Reset the Setup menu to factory defaults.
NO
= not reset
YES
= reset
Filter [FiltEr]
Set the amount of signal filtering.
LOW
= less stability, faster stabilization time
MED
= normal stability, stabilization time
HI
= greater stability, slower stabilization time
AZT [AZt]
Set the automatic zero tracking functionality.
OFF
= disabled
0.5d
= the display will maintain zero until a change of 0.5 divisions per second has
been exceeded.
1d
= the display will maintain zero until a change of 1 divisions per second has been
exceeded.
3d
= the display will maintain zero until a change of 3 divisions per second has been
exceeded.
Stable Range [Stable]
Set the amount the reading can vary while the stability symbol remains on.
0.5d
= 0.5 balance division
1d
= 1 balance division
2d
= 2 balance division
5d
= 5 balance division
no, yes
Low, Med, High
off, 0.5d, 1d, 3d
0.5d, 1d, 2d, 5d
off, on, auto
off, on, on-acc
off, 1, 5, 10
Exit menu