Air Cooling Kit Installation; Gas Purge Valve Installation - Varian 969-9369 Manual De Instrucciones

Ocultar thumbs Ver también para 969-9369:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 12

AIR COOLING KIT INSTALLATION

A fan kit model 969-9314 is available for air cooling
the pump.
Fan specification:
− air flow: 59 l/s (120 CFM)
− input voltage: 120 Vac 50-60 Hz
− maximum power: 17 W
− dimensions: 127 x 127 mm (5 x 5 inches)
Position the fan on the pump, ensuring that the
holes in the plate line up with those in the pump
base.
Insert the respective washers and screws and
tighten with a screwdriver.
Connect the fan to the controller.
All manuals and user guides at all-guides.com
30
TECHNICAL INFORMATION

GAS PURGE VALVE INSTALLATION

A gas purge valve is available to protect the pump
bearings against particulate and corrosive gases
that could move into the pump.
To install the gas purge valve it is necessary to un-
screw the purge port cover as shown in the follow-
ing figure,
87-900-891-01(C)
Tabla de contenido
loading

Este manual también es adecuado para:

969-9370969-9371969-9372

Tabla de contenido