Molybdenum Hr With Liquid Reagent - Lovibond MD100 Instrucciones

Ocultar thumbs Ver también para MD100:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 196
Methods
GB
Mo 2
0.0.0
Zero
!
Test
Mo 2
RESULT
86

Molybdenum HR with Liquid reagent

0.6 - 60 mg/l Mo
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Fill the vial with drops of the same size by holding the bottle vertically
and squeeze slowly:
10 drops KS63 (Thioglycolate)
Close the vial tightly with the cap and swirl several times to mix the
contents.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 5 minutes.
(Countdown can be activated, see page 57)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display as Molybdenum.
marks
MD100_7 12/2016
Tabla de contenido
loading

Tabla de contenido