Esc
<Calibration>
511 Fluorescein 2P
T1: 0 ppb
press TEST
Test
T1 accepted
T2: (010 ... 300): 100
or <shift> + <x>
Test
<Calibration>
511 Fluorescein 2P
Calibration accepted
66
Confirm with [ ] key.
It is possible to cancel the entry by [ESC].
The display shows:
1. Fill a clean vial (24 mm Ø) with 10 ml of deionised wa-
ter or 0 ppb Fluorescein standard solution (blank),
close tightly with a black cap.
2. Press TEST key.
3. The display shows:
Confirm (e.g. 100) with [ ] key key to store the predeter-
mined standard concentration or enter a concentration
in the range from 10 to 300, e.g.: [Shift] + [1][0][0]
Confirm with [ ] key.
4. Remove the vial from the sample chamber.
5. Perform a measurement with the standard of the set
concentration. Place the vial in the sample chamber
making sure that the
6. Press TEST key.
7. The display shows:
8. Confirm with [ ] key.
The calibration is saved.
marks are aligned.
07/2017