Lovibond Tintometer Manual Del Usuario página 65

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 128
Test
<Calibration>
501 PTSA 2P
Calibration accepted
Fluorescein 2P Method 511
Do not calibrate in direct sunlight!
4
Mode
<Calibration>
>> 191 Ca-hardness 2
191 reset 0 cali.
170 Fluoride L
500 PTSA
501 PTSA 2P
510 Fluorescein
511 Fluorescein 2P
07/2017
Confirm with [ ] key.
4. Remove the vial from the sample chamber.
5. Perform a measurement with the standard of the set
concentration. Place the vial in the sample chamber
making sure that the
6. Press TEST key.
7. The display shows:
8. Confirm with [ ] key.
The calibration is saved.
0
Press [MODE], [Shift] + [4] [0] keys.
Confirm with [ ] key.
The display shows:
Press [ESC] key to return to mode selection.
Select the desired determination (511 Fluorescein 2P) with
the arrow keys [ ] and [ ].
marks are aligned.
65