Metrohm 798 MPT Titrino Instrucciones De Uso página 80

Ocultar thumbs Ver también para 798 MPT Titrino:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

3.1. General rules
E196
E197
E198
E199
E205
E212
E213
E214
E270
76
Result is out of limits.
Exit: The error message disappears on next start or on recalculation.
Sample size is out of limits.
Exit: The error message disappears on next start or on introduction of new
sample size.
Validation interval is expired.
Exit: The error message disappears on next start or clear counter with
&Config.Monitoring.Validation.ClearCount $G.
Service date is reached.
Exit: The error message disappears on next start or change date in
&Config.Monitoring.Service.Date.
Calibration interval is expired.
Exit: The error message disappears on next calibration or if you delete the
calibration.
Transmission error from Remote Box. Unknown characters.
Exit: Rectify error and switch Titrino off and on again.
Time-out error from PC keyboard (Remote Box)
Exit: Rectify error and switch Titrino off and on again.
Check Remote Box. Remote Box not (properly) connected but activated in
&Config.Periph.RemoteBox.
Exit: Rectify error and switch Titrino off and on again.
Overload in dosing element.
Exit: <STOP> then <CLEAR>. The dosing element will be initialized. If the
error reappears, check the wet part.
798 MPT Titrino
Tabla de contenido
loading

Tabla de contenido