3.2. Remote control commands
120
L6
Activate pulse + can be set in TIP
L7
Buret volume pulses
L8
Remote box active
L12
Sample size out of limits
L13
Result out of limits
.ResetLines: Lines are set to the inactive status (= high).
3.2.2.114. Assembly.Stirrer
Switching stirrer ON/OFF.
3.2.2.115. Setup.Comport
Selects the Titrino COM for the output of automatic info:
&Setup.Keycode
&Setup.Trace
&Setup.SendMeas
&Setup.AutoInfo
3.2.2.116. Setup.Keycode
ON means the key code of a key pressed on the Titrino is outputted. The key
code comprises 2 ASCII characters; table of the keys with their code, see
page 140. A keystroke of key 11 is sent as follows:
#11
The beginning of the message is marked by a space (ASCII 32).
3.2.2.117. Setup.Tree.Short
Setup.Tree.ChangedOnly
Definition of the type of answer to $Q.
.Short:
With "ON", each path is sent with only the necessary amount of
characters in order to be unequivocal (printed in bold in this manual). A com-
bination of .Short and .ChangedOnly is not possible.
.ChangedOnly:
Sends only the changed values, i.e. values which have
been edited. All paths are sent absolute, i.e. from the root.
3.2.2.118. Setup.Trace
The Titrino automatically reports when a value has been confirmed with <en-
ter> at the Titrino. Message, e.g.:
&SmplData.OFFSilo.Id1"Trace"
The beginning of the message is marked by a space (ASCII 32).
3.2.2.119. Setup.Lock.Keyboard
Setup.Lock.Config
Setup.Lock.Parameter
Setup.Lock.SmplData
Setup.Lock.UserMeth.Recall
Setup.Lock.UserMeth.Store
Setup.Lock.UserMeth.Delete
ON, OFF
1, 2, 1&2
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
798 MPT Titrino