3.1.4 Error messages

E8
E9
E10
E18
E20
E21
E22
E23
E24
E26
E27
E28
E29
E30
E31
E32
E33
E34
798 MPT Titrino
Error messages are added to the status messages and separated from them
by the sign ";".
Card read/write error.
Exit: Send new command.
Wrong card, a card has been removed/inserted during the inquiry.
Exit: Send new command.
The card has lost data.
Exit: Send new command.
Card battery low (it is between 2.37...2.64 V).
Exit: Send new command.
Check exchange unit.
Exit: Mount Exchange Unit (properly) or &m $S.
Check electrode, short circuit.
Exit: Rectify fault or &m $S.
Check electrode, break.
Exit: Rectify fault or &m $S.
Division by zero.
Exit: The error message disappears on next startup or on recalculation.
Check drive unit.
Exit: Connect drive unit (correctly) or &m $S.
Manual stop.
Exit: The error message disappears on next startup.
Stop V reached in SET.
Exit: The error message disappears on next startup.
Wrong object call up
Exit: Send correct path for object. Start path at root.
Wrong value or no value allowed.
Exit: Send correct value or call up new object.
Wrong trigger, this trigger is not allowed or carrying-out of action not
possible.
Exit: Send correct trigger (exception: $D) or call up new object.
Command is not possible in active status. Repeat command in inactive
status.
Exit: Send new command.
Command is not possible during titration. Repeat command during the
conditioning phase or in inactive status.
Exit: Send new command.
Value has been corrected automatically.
Exit: Send new command.
Instrument at the end of the titration and sample data is edited; the instrument
at rest or editing during filling.
Exit: &m $S.
3.1 General rules
73