Guidance and manufacturer's declaration – electromagnetic immunity
Electrical fast tran-
sient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Conducted RF
IEC 61000-4-6
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE:
UT is the a.c. mains voltage prior to application of the test level.
± 2 kV @ 100 kHz repeti-
tion frequency for power
supply lines
± 2 kV @ 100 kHz repeti-
tion frequency for input/
output lines
Power inputs
± 0,5 kV, ± 1 kV Line-
to-Line
± 0,5 kV, ± 1 kV, ± 2 kV
Line-to-Ground
Signal input/outputs:
± 2 kV Line-to-Ground
0,15 MHz – 80 MHz
3V, 80 % AM at 1 kHz
ISM bands between 0,15
MHz and 80 MHz
6V, 80 % AM at 1 kHz
Voltage Dips:
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
phase angles
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
Voltage interruptions:
0 % UT; 250/300 cycle
30 A/m
50 Hz or 60 Hz
± 2 kV @ 100 kHz
repetition frequency for
power supply lines
± 2 kV @ 100 kHz
repetition frequency for
input/output lines
Power inputs
± 0,5 kV, ± 1 kV Line-
to-Line
± 0,5 kV, ± 1 kV, ± 2 kV
Line-to-Ground
Signal input/outputs:
± 2 kV Line-to-Ground
0,15 MHz – 80 MHz
3V, 80 % AM at 1 kHz
ISM bands between
0,15 MHz and 80 MHz
6V, 80 % AM at 1 kHz
Voltage Dips:
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315° phase angles
0 % UT; 1 cycle
Single phase: at 0°
70 % UT; 25/30 cycles
Single phase: at 0°
Voltage interruptions:
0 % UT; 250/300 cycle
30 A/m
50 Hz or 60 Hz
8
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the
user of the ARC-EPMS
unit requires continued
operation during power
mains interruptions, it is
recommended that the
ARC-EPMS unit be
powered from an uninter-
ruptible power supply or
a battery.
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.