STANDARD AND SLIM ELECTRONIC TORQUE-ANGLE BLUETOOTH
AND ELECTRONIC TORQUE-ANGLE BLUETOOTH
STD
Note: Clock Date and Time must be set and Cal interval enabled before calibration countdown
will function (see Setting Date and Time Setting Calibration Interval).
Note: Calibration Countdown is number of days before wrench is locked out from use following
an expired Cal interval. Maximum count is 10 days. Countdown number of days before lockout
CAL NEEDED
1 DAY
torque while countdown is remaining, torque or angle is immediately displayed. Red LEDs and
buzzer turn on if countdown has expired. Pressing ENTER button after countdown has expired
will display Password Entry screen.
Note: An entered Cal Countdown value of 0 days will disable calibration countdown and if Cal
interval is enabled, wrench does not lock when CAL NEEDED" message is displayed.
Note: Calibration Countdown and Calibration Cycles can be enabled simultaneously. Both
number of days and number of cycles before wrench is locked are displayed:
CAL NEEDED
10 DAYS
100 CYCLES
SETTING DATE AND TIME
Set Date/Time function allows user to set real-time-clock date and time for time stamping data records,
recording last calibration date and notifying user of an expired calibration interval.
®
WRENCH
®
SCREWDRIVER
SLIM
CAL NEEDED
1 DAY
CAL NEEDED
10 DAYS
100 CYCLES
43