General Setup

h
You can configure the M400 PA transmitter via configuration program or via display
on-site.
For step 1 to 4 see documentation of the configuration program.
Prerequisite: The M400 PA transmitter and the sensor are mounted and electrically
connected.
Note: The GSD file is on the supplied CD-ROM.
1. Copy the GSD file into the GSD-directory of the configuration program.
GSD file: METT0E8A.gsd
2. Update hardware catalogue.
3. Integrate M400 PA transmitter into the DP master system.
4. Set PROFIBUS address for M400.
Factory setting: 126, Input range: 0 ...125
5. Perform parameterization. For further information see documentation of the
M400 PA on the supplied CD-ROM and of the configuration program.